6.8 Auto Star t Function
KW-2000 incorporates Auto Start Function.
This function starts measuring automatically when alignment meets the measurement requirement, and
also prints out automatically when the measurement of both eyes completes.
(1) When you wish to measure with auto start function,
select 'AUTO3' or 'AUTO5' in 'START' of the Menu
screen
(2) Then, press switch to go back to the
measurement screen.
(3) Auto Start mode is activated when 'AUTO' and arrows
below 'AUTO' appear on the top of the screen (see on
the left).
You can start measurement from either right or left eye.
(4) Carry out alignment. Measurement will automatically
start when alignment is achieved.
When the measurement of one eye finishes, an arrow on
either left or right side you have measured will
disappear. In addition, an arrow to urge the shift to the
other eye will appear and flash in the center of the
screen.
(5) Measure the other eye in the same manner.
The operator shifts the body of the unit to the
eye which has not measured according to the
arrow.
When the body is shifted, the arrow flashing in
the center of the screen will disappear.
NOTE
START MANUAL AUTO3 AUTO5
REF NORMAL QUICK( 3)
STEP 0.25 0.12
VD(mm) 0 10 12 13.5 15
CYL
KERATO RADIUS DIOPT
PRINT FORM ALL ECONO OFF
PRINT OUT TOP BOTTOM
DATA SCREEN ON OFF
W-D(cm) OFF 30 35 40 45
TARGET LIGHT BRIGHT NORMAL DARK
SCREEN ADJ.
Ker.Ref.
(
0 )
(
0 )
R
R
K
MENU
Ker.Ref.
(
0 )
(
0 )
R
R
K
MENU
The arrow to urge the
shift flashes.
When Auto Start Function is activated, measurement is always taken according
to the setting of START regardless of REF setting in the menu screen.
NOTE
For alignment, refer to '6.2 Alignment'.