VXI 3002 Portable Generator User Manual


 
5-8
(4) Measure the level of the second and third harmonics on the spectrum analyzer at each of the
carrier frequencies shown in Table 5-15 checking that the results are within specification.
(5) Set the UUT RF level to +7 dBm and repeat (4) using Table 5-16.
(6) Set the UUT RF level to +13 dBm and repeat (4) using Table 5-17.
(7) Set the UUT RF level to +19 dBm and repeat (4) up to 1.2 GHz using Table 5-18.
Non-harmonics
Test procedure
(1) Press CAL on the spectrum analyzer.
(2) Connect the test equipment as shown in Fig. 5-4.
(3) On the UUT set:
Carr Freq 1201 MHz
RF Level 0 dB
(4) Measure the level of the non-harmonics on the spectrum analyzer at each of the carrier
frequencies shown in Table 5-19 checking that the results are within specification.
Residual FM
Test procedure
UUT
C2562
RF OUT
RF INPUT
Measuring
receiver
Fig. 5-5 Residual FM test set-up
(1) Connect the test equipment as shown in Fig. 5-5.
(2) On the UUT set:
Carr Freq 1 [GHz]
RF Level 0 dBm
(3) On the measuring receiver select FM, 300 Hz high-pass filter, 3.4 kHz low-pass filter and
enable averaging.
(4) Measure the residual FM checking that the result is within the specification shown in
Table 5-20.