VXI 3002 Portable Generator User Manual


 
3-9
Output control
These commands allow you to download and store settings without the output changing.
OUTPUT
[not used alone]
:DISABLE
Allows user to download and store settings in the normal way without
the output of the instrument changing until the OUTPUT:ENABLE
command is received.
:ENABLE
Enables the instrument outputs such that the outputs will adjust to the
values specified by commands sent while the outputs were disabled.
Note: It is up to the user to ensure that the last command sent, prior to
OUTPUT:ENABLE, is such that the RF output is set to a safe
level.
Data type :
None
Allowed suffices :
None
Default suffix :
None
Examples: OUTPUT:DISABLE
CFRQ 300MHZ; RFLV 10DBM; MODE AM; AM 40PCT;
AM:ON; MOD:ON; STO 200
CFRQ 400MHZ; RFLV 7DBM; STO 201
CFRQ 500MHZ; RFLV 5DBM; STO 202
CFRQ 600MHZ; RFLV 4DBM; STO 203
OUTPUT;ENABLE
RCL 200
RCL 201
RCL 202
RCL 203
OUTPUT?
Prepares message containing information on output control setting in
the following format:
: OUTPUT: <status>
where: <status> is a program mnemonic indicating whether the
output control is ENABLED or DISABLED
Examples: :OUTPUT:ENABLE
:OUTPUT:DISABLE