VXI 3002 Portable Generator User Manual


 
3-22
Memory - triggering
For external triggering methods, refer to 'Trigger source' at the end of this section.
MTRIG
:ON
Enables memory recall triggering to be activated by *TRG command or
by external triggering
:OFF
Disable memory recall triggering
Data type :
None
Allowed suffices :
None
Default suffix :
None
Examples: MTRIG:ON
MTRIG:OFF
MTRIG?
Prepares message containing information on memory triggering state in the
following format:
:MTRIG:<status>
where: <status> is a program mnemonic indicating whether the
memory recall triggering is enabled (:ON) or disabled (:OFF)
Example: :MTRIG:ON
Memory - protection
These commands enable you to either write protect a block of stores (or a single store) to prevent
accidental overwriting or to unprotect it. For a single store set both start and stop numbers the
same. Note that any protection applied to RAM will be lost once the instrument has been switched
off.
MPROT
[not used alone]
:START
Set the start of the memory block which is to be protected/unprotected
:STOP
Set the end of the memory block which is to be protected/unprotected
Data type :
Decimal Numeric Program Data
Allowed suffices :
None
Default suffix :
None
:ON
Set memory protection ON for the selected memory block
:OFF
Set memory protection OFF (i.e. unprotected) for the selected memory
block
Data type :
None
Allowed suffices :
None
Default suffix :
None
Examples: MPROT:START 100;STOP 150
MPROT:ON