VXI 3002 Portable Generator User Manual


 
3-8
:UNITS
Select default RF level units.
Data type : Character Program Data (DBM, DBV, DBMV, DBUV, V, MV or UV)
Allowed suffices :
None
Default suffix :
None
Examples: RFLV:VALUE -27.3DBM;ON
RFLV:TYPE PD;VALUE 1.23UV
:LIMIT
Set RF Level max limit (short form)
:VALUE Set RF Level max limit
Data type :
Decimal Numeric Program Data
Allowed suffices :
Any one of: DBM, DBV, DBMV, DBUV, V, MV or UV
Default suffix :
DBM unless changed by UNITS command
:ENABLE Enable limit
:DISABLE Disable limit
:OFFS
Set RF Level offset for given frequency band (short form)
:VALUE Set RF Level offset for given frequency band
Data type : Decimal Numeric Program Data
Allowed suffices :
DB only
Default suffix :
DB
:ENABLE Enable offsets
:DISABLE Disable offsets
:SAVE Save offsets in non-volatile memory
RFLV?
Prepares message containing information on RF Level setting in the
following format:
:RFLV:UNITS <unit>;TYPE <type>;VALUE <nr2>;INC <nr2>;<status>
where: <unit> is character program data defining the default RF
level units (DBM, DBV, DBMV, DBUV, V, MV or UV),
<type> is character program data indicating EMF or PD and
<status> is a program mnemonic indicating whether the RF
output is ON or OFF
Examples: :RFLV:UNITS DBM;TYPE PD;VALUE -103.5;INC 2.0;ON
:RFLV:UNITS DBV;TYPE EMF;VALUE -83.2;INC 0.5;ON
RFLV:LIMIT?
Prepares message containing information on RF Level max limit setting
in the following format:
:RFLV:LIMIT:VALUE<nr2>;<status>;
Examples: :RFLV:LIMIT:VALUE-20.0;ENABLE
RFLV:OFFS?
Prepares message containing information on RF Level offset in the
following format:
:RFLV:OFFS:VALUE<nr2>;<status>;
Examples: :RFLV:OFFS:VALUE-3.2;ENABLE