Pentax R-315(N) Telescope User Manual


 
7-3 Changing the Atmospheric Pressure
The atmospheric pressure setting can be changed only when Atmospheric
Correctionhas been set to ATM INPUTin Initial Setting 1.
Example: Setting the pressure to 900 hPa
Press [F4] [CORR] in mode B.
(If the instrument is in mode A, press [F5]
[MODE] to enter mode B.)
Press [F4] [h] to move the cursor to
4.PRESSand press the [F5] [SELECT] to
enable the temperature to be changed.
Clear the exiting values by pressing [CLEAR]
key.
Input 900 by pressing the numeric keys.
50
7 CORRECTION MODE
SELECT
CORRECTION
1.
PRISM CONST
0mm
2.
SHEET CONST
0mm
3.TEMP : +15°C
4.PRESS : 1013hPa
5.ppm 0ppm
CORRECTION
1.
PRISM CONST
0mm
2.
SHEET CONST
0mm
3.TEMP : +15°C
4.PRESS : 1013hPa
5.ppm 0ppm
SELECT
CORRECTION
1.
PRISM CONST
0mm
2.
SHEET CONST
0mm
3.TEMP : +15°C
4.PRESS : 0900hPa
5.ppm 0ppm
CLEAR
PENTAX R-300 (p20-52) 10/30/02 11:08 AM Page 50