SECTION 1 GENERAL
1-10
Other functions Relative value display Carrier frequency and output level
Offset display Carrier frequency and output level
Memory 1000 panel setting conditions can be stored and recalled.
Recall mode: All panel settings, Frequency only, Frequency and Level.
Trigger function An external trigger signal (input from the BNC connector on the rear
panel, TTL level) can be used to execute a pre-programmed operation
sequence (except for operation of the power switch, Preset key, Local
key, and rotary knob).
Max. number of sequence steps of trigger program: 20 steps
Backup When switched on, the generator restores the same setting conditions
that existed immediately before it was last powered off.
However, the following are not restored:
• Data which was in the middle of entry
• Remote status
• Data which was in the middle of GPIB transfer
• Operating status of RPP
GPIB Can control all functions, except for pre-programmed operations
controlled by the trigger function, and operation of the power switch,
Local key, rotary knob, and knob resolution set keys.
Interface: SH1, AH1, T5, L3, TE0, SR1, RL1, PP0, DC1, DT1, C0,
E2
Reverse power Maximum reverse input 50 W (≤1040 MHz), 25 W (1040 to 2080 MHz, MG3642A only)
protection power ±50 Vac
General Power Voltage: *Vac (Up to max. 250 V)
Frequency: 47.5 to 63 Hz, 380 to 420 Hz
Capacity: 200 VAmax
Environmental performance Working temperature range: 0 to 50°C
Storage temperature range: –30 to 71°C
Conducted disturbance: EN 61326-1: 2006 (Class A)
Radiated disturbance: EN 61326-1: 2006 (Class A)
Harmonic Current Emission: EN 61000-3-2: 2006 (Class A)
Electrostatic Discharge: EN 61326-1: 2006 (Table 2)
Electromagnetic Field Immunity: EN 61326-1: 2006 (Table 2)
Fast Transient / Burst: EN 61326-1: 2006 (Table 2)
Surge: EN 61326-1: 2006 (Table 2)
Conducted RF: EN 61326-1: 2006 (Table 2)
Power Frequency Magnetic Field: EN 61326-1: 2006 (Table 2)
Voltage Dips / Short Interruptions: EN 61326-1: 2006 (Table 2)
Dimensions and mass 177 H×320 W×451 Dmm, ≤20 kg
* Please specify a nominal voltage in the range from 100 and 240 V when ordering the product.
1-10