Anritsu MG3642A Portable Generator User Manual


 
SECTION 7 PERFORMANCE TEST
7-9
(4) Test procedure
STEP PROCEDURE
FM deviation
1. Set the MG3641A/MG3642A output level to +7 dBm.
2. Set the MG3641A/MG3642A frequency and modulation analyzer receive frequency to the
expected frequency.
3. Turn on the MG3641A/MG3642A FM, and set Source to Int1 1 kHz.
4. Set the MG3641A/MG3642A FM frequency deviation.
5. Read the indicator of the modulation analyzer.
6. Change the MG3641A/MG3642A FM frequency deviation and repeat the measurement.
FM distortion
7. Set the MG3641A/MG3642A FM frequency deviation to 3.5 kHz or 22.5 kHz.
8. Measure the demodulation output of the modulation analyzer using the distortion meter.
(5) Note on test
The demodulation zone of the modulation analyzer must be within 0.3 to 3 kHz at measurement of the FM
deviation and 0.3 to 15 kHz or 0.3 to 20 kHz at measurement of distortion rate.
Using a modulation analyzer with much residual FM affects the measurement of distortion if the FM deviation
rate is lower. Use a modulation analyzer with a little residual FM.