Agilent Technologies 66311B Portable Generator User Manual


 
Language Dictionary - 8
99
Common Commands
Common commands begin with an * and consist of three letters (command) or three letters and a ?
(query). They are defined by the IEEE 488.2 standard to perform common interface functions. Common
commands and queries are categorized under System, Status, or Trigger functions and are listed at the
end of the chapter. The dc source responds to the following common commands:
Table 8-2. Common Commands Syntax
*CLS
Clear status
*ESE <n>
Standard event status enable
*ESE?
Return standard event status enable
*ESR?
Return event status register
*IDN?
Return instrument identification
*OPC
Enable "operation complete" bit in ESR
*OPC?
Return a "1" when operation complete
*OPT?
Return option number
*PSC <bool>
Power-on status clear state set/reset
*PSC?
Return power-on status clear state
*RCL <n>
Recall instrument state
*RST
Reset
*SAV <n>
Save instrument state
*SRE <n>
Set service request enable register
*SRE?
Return service request enable register
*STB?
Return status byte
*TRG
Trigger
*TST?
Perform selftest, then return result
*WAI
Hold off bus until all device commands done
Programming Parameters
The following table lists the output programming parameters.
Table 8-3. Output Programming Parameters
Parameter Value
[SOUR:]CURR[:LEV][:IMM] MAX and
[SOUR:]CURR[:LEV]:TRIG MAX
3.0712 A
[SOUR:]CURR2[:LEV][:IMM] MAX and
[SOUR:]CURR2[:LEV]:TRIG MAX
1.52 A
*RST Current Value 10% of MAX value
[SOUR:]VOLT[:LEV][:IMM]MAX and
[SOUR:]VOLT[:LEV]:TRIG MAX
15.535 V
[SOUR:]VOLT2[:LEV][:IMM]MAX and
[SOUR:]VOLT2[:LEV]:TRIG MAX
12.25 V
*RST Voltage Value 0 V
[SOUR:]VOLT:PROT[:LEV] MAX 22 V
*RST OVP Value MAX
OUTP:PROT:DEL MAX 2,147,483.647
*RST Protection Delay Value 0.08 seconds
SENS:CURR:RANG (not valid for Agilent
66111A)
Low range = 0 20 mA
High Range = 20 mA MAX
*RST Current Range Value MAX