Agilent Technologies 66311B Portable Generator User Manual


 
Language Dictionary - 8
133
TRIGger:SEQuence2:SLOPe:DVM
TRIGger:ACQuire:SLOPe:DVM
Agilent 66311D/66309D only
This command sets the slope of the DVM input signal.
POSitive
measurement triggering occurs on the rising edge.
NEGative
measurement triggering occurs on the falling edge.
EITHer
measurement triggering occurs on either edge.
Command Syntax
TRIGger:SEQuence2:SLOPe:DVM<slope>
TRIGger:ACQuire:SLOPe:DVM<slope>
Parameters
EITHer | POSitive | NEGative
*RST Value
POSitive
Examples
TRIG:SEQ2:SLOP:DVM POS TRIG:ACQ:SLOP:DVM EITH
Query Syntax
TRIGger:SEQuence2:SLOPe:DVM?
TRIGger:ACQuire:SLOPe:DVM?
Returned Parameters
<CRD>
Related Commands
TRIG:SEQ2:LEV:DVM
TRIGger:SEQuence2:SLOPe:VOLTage
TRIGger:ACQuire:SLOPe:VOLTage
This command sets the slope of an internally triggered voltage measurement.
POSitive
triggering occurs on the rising edge.
NEGative
triggering occurs on the falling edge.
EITHer
triggering occurs on either edge.
Command Syntax
TRIGger:SEQuence2:SLOPe:VOLTage<slope>
TRIGger:ACQuire:SLOPe:VOLTage<slope>
Parameters
EITHer | POSitive | NEGative
*RST Value
POSitive
Examples
TRIG:SEQ2:SLOP:VOLT POS TRIG:ACQ:SLOP:VOLT EITH
Query Syntax
TRIGger:SEQuence2:SLOPe:VOLTage?
TRIGger:ACQuire:SLOPe:VOLTage?
Returned Parameters
<CRD>
Related Commands
TRIG:SEQ2:SLOP:CURR