Agilent Technologies 66311B Portable Generator User Manual


 
8 – Language Dictionary
120
[SOURce:]VOLTage2
Agilent 66309B/D only
This command sets the output voltage level of the auxiliary output.
Command Syntax
[SOURce:]VOLTage2[:LEVel][:IMMediate][:AMPLitude]<NRf+>
Parameters
see Table A-3
Default Suffix
V (volts)
*RST Value
0
Examples
VOLT2 500mV !set output2 voltage to 0.5V
Query Syntax
[SOURce:]VOLTage2[:LEVel][:IMMediate][:AMPLitude]?
Returned Parameters
<NR3>
[SOURce:]VOLTage:PROTection
This command sets the overvoltage protection (OVP) level of the dc source. If the output voltage exceeds
the OVP level, then the dc source output is disabled and the Questionable Condition status register OV
bit is set (see chapter 7 about programming the status registers). An overvoltage condition can be cleared
with the OUTP:PROT:CLE command after the condition that caused the OVP trip is removed. The OVP
always trips with zero delay and is unaffected by the OUTP:PROT:DEL command. The OVP function
can be disabled using the VOLT:PROT:STAT command.
Command Syntax
[SOURce:]VOLTage:PROTection[:LEVel]<NRf+>
Parameters
see Table 8-3
Default Suffix
V (volts)
*RST Value
MAX
Examples
VOLT:PROT 21 !set voltage protect to 21V
Query Syntax
[SOURce:]VOLTage:PROTection[:LEVel]?
Returned Parameters
<NR3>
Related Commands
OUTP:PROT:CLE VOLT:PROT:STAT
[SOURce:]VOLTage:PROTection:STATe
This command enables or disables the overvoltage protection (OVP) function.
CAUTION: Disabling the overvoltage protection function may cause excessive output voltages, such
as can occur if remote sense leads are disconnected, to damage the equipment under test.
Command Syntax
[SOURce:]VOLTage:PROTection:STATe <bool>
Parameters
0 | 1 | OFF | ON
*RST Value
OFF
Examples
VOLT:PROT:STAT 0 !voltage protection is OFF
VOLT:PROT:STAT 1 !voltage protection is ON
Query Syntax
[SOURce:]VOLTage:PROTection:STATe?
Returned Parameters
<NR1>0 or 1
Related Commands
OUTP:PROT:CLE *RST VOLT:PROT