SENSe:FUNCtion
This command configures the sensing function for triggered measurements. The dc source has up to three
measurement sensors as described below. The query returns the function setting.
CURRent
Senses the current measurement at the main output (output 1)
DVM
Senses the voltage measurement at the DVM inputs (Agilent 66311D/66309D
only)
VOLTage
Senses the voltage measurement at the main output (output 1)
Command Syntax
SENSe:FUNCtion <function>
Parameters
"VOLTage" | "CURRent" | "DVM"
Examples
SENS:FUNC "VOLT"
Query Syntax
SENSe:FUNCtion?
Returned Parameters
<SRD>
SENSe:PROTection:STATe
This command enables or disables the open sense lead detection. The default setting for this function is
disabled. To permanently enable open sense lead detection, program this command on, save this state in
location 0 using *SAV 0, and set the output power-on state to recall state 0 (OUTP:PON:STAT RCL 0).
Command Syntax
SENSe:PROTection:STATe <bool>
Parameters
0 | OFF | 1 | ON
*RST Value
0
Examples
SENS:PROT:STAT ON
Query Syntax
SENSe:PROTect:STATe?
Returned Parameters
<NR3>
SENSe:SWEep:OFFSet:POINts
This command defines the offset in a data sweep when an acquire trigger is used. Negative values
represent data samples taken prior to the trigger. Positive values represent the delay after the trigger
occurs but before the samples are acquired.
Command Syntax
SENSe:SWEep:OFFSet:POINts <NRf+>
Parameters
-4095 through 2,000,000,000
*RST Value
0
Examples
SENS:SWE:OFFS:POIN -2047
Query Syntax
SENSe:SWEep:OFFSet:POINts?
Returned Parameters
<NR3>
Related Commands
SENS:SWE:TINT SENS:SWE:POIN MEAS:ARR
SENSe:SWEep:POINts
This command defines the number of points in a measurement.
Command Syntax
SENSe:SWEep:POINts<NRf+>
Parameters
1 through 4096
*RST Value
2048
Examples
SENS:SWE:POIN 1024
Query Syntax
SENSe:SWEep:POINts?
Returned Parameters
<NR3>
Related Commands
SENS:SWE:TINT SENS:SWE:OFFS MEAS:ARR