Agilent Technologies 66311B Portable Generator User Manual


 
Verification and Calibration - B
153
Table B-4. Verification Test Record for Agilent Model 66309B/D
Model Agilent___________ Report No.____________ Date_____________
Test Description Minimum
Specification
Recorded Results Maximum
Specification
Voltage Programming and Measurement
Low Voltage Vlo
10 mV
_____mV +10 mV
Front Panel measurement Vfplo
Vlo 5 mV
_____mV Vlo +5 mV
High Voltage Vhi 14.982 V _______V 15.018 V
Front Panel measurement Vfphi
Vhi 9.5 mV
_______V Vhi +9.5 mV
Outp 2 Low Voltage V2lo
40 mV
_____mV +40 mV
Outp 2 Front Panel measurement V2fplo
V2lo 15 mV
_____mV V2lo +15 mV
Outp 2 High Voltage V2hi 11.936 V _______V 12.064 V
Outp 2 Front Panel measurement V2fphi
V2hi 39 mV
_______V V2hi +39 mV
Current Programming and Measurement (High Range)
Low Current Ilo
1.33 mA
_____mA 1.33 mA
High Current Ihi 2.9972 A _______A 3.0028 A
Front Panel measurement Ifphi
Ihi 6.5 mA
_______A Ihi +6.5 mA
Outp 2 Low Current I2lo
4.5 mA
_____mA 4.5 mA
Outp 2 High Current I2hi 1.492 A _______A 1.508 A
Outp 2 Front Panel measurement I2fphi
I2hi 6 mA
_______A I2hi +6 mA
Current Measurement (Low Range)
Low Current measurement Ilrlo
Ilrlo 2.5 µA _____µA Ilrlo +2.5 µA
High Current measurement Ilrhi
Ilrhi 22.5 µA
_____mA
Ilrhi +22.5 µA
Current Sink Measurement
Low Current Sink measurement Isinklo
Isinklo 22.5 µA
_____mA
Isinklo +22.5 µA
High Current Sink measurement Isinkhi
Isinkhi 5.1 mA
_____A Isinkhi +5.1 mA
DVM Voltage Measurement (Agilent 66309D only)
Positive Voltage measurement Dvpos
Dvpos 15 mV
_____V Dvpos +15 mV
Negative Voltage measurement Dvneg
Dvneg 6.8 mV
_____V Dvneg +6.8 mV