When an abnormal condition is detected during op
-
eration, the 681XXC displays an error message to
indicate that the output is abnormal or that a signal
input or data entry is invalid. It also displays warn
-
ing messages to alert the operator to conditions that
could cause an inaccurate signal generator output.
Status messages to remind the operator of current
menu selections or settings are also generated.
Table 6-2 is a summary list of possible error mes
-
sages that can be displayed during normal opera
-
tions. Table 6-3 is a summary list of possible
warning/status messages.
6-8 681XXC OM
OPERATOR ERROR AND WARNING/
MAINTENANCE STATUS MESSAGES
Error Message Description
ERROR
Displayed (on the frequency mode title bar) when (1) the
output frequency is not phase-locked or (2) an invalid
entry causes a frequency range error.
LOCK ERROR
Displayed (in the frequency parameters area) when the
output frequency is not phase-locked. The frequency
accuracy and stability of the RF output is greatly reduced.
Normally caused by an internal component failure. Run
self-test to verify malfunction.
RANGE
Displayed (in the frequency parameters area) when (1)
the analog sweep start frequency entered is greater than
the stop frequency, (2) the dF value entered results in a
sweep outside the range of the instrument, (3) the step
size value entered is greater than the sweep range, (4)
the number of steps entered results in a step size of less
than 1 kHz (0.1 Hz with Option 11) or 0.1 dB (0.001 mV),
or (5) the step sweep time entered divided by the number
of steps entered does not result in a dwell time of
<10 ms. Entering valid values usually clears the error.
ERR
Displayed (in the modulation status area) when either the
external AM modulating signal or the external FM
modulating signal exceeds the input voltage range. In ad
-
dition, the message Reduce AM (FM) Input Levelap
-
pears at the bottom of the AM (FM) status display. AM
(FM) will be turned off until the modulating signal is in the
input voltage range.