Agilent Technologies 75000 Series C Portable Generator User Manual


 
DC Adjustment Procedure (cont’d)
Test Procedure
(cont’d)
5. Trigger the DMM and note the reading.
6. Send the reading to the AFG:
CAL:DC:POINT? <reading>
where <reading> is the DMM reading from step 5. The AFG will
return, in order, the number of the current cal point and an error code.
Any non-zero error code indicates a failure.
Table 3-1. DMM Range Setting for Cal Points 41 and 43
Absolute Value
of DC Offset DMM Range
Less than 10 mV
100
Between 10 mV and 100 mV
1000
Greater than 100 mV Offset too high - abort cal
Table 3-2. DC Calibration Points
Cal
Point
DMM Settings
(changes only)
1
29
31
33
41
43
DCV, 10 V range, 10 NPLC
1 V range, 20 NPLC
10 V*
100 mV range
4-wire ohms, offset comp on*
4-wire ohms, offset comp on*
*See Step 4 of the "DC Adjustment Procedure" for
special instructions.
Agilent E1445A Service Manual Adjustments 89