Test 2-5: AC Flatness - 250 kHz Filter
Description
The purpose of this test is to verify that the AFG meets its specifications for
AC flatness with the 250 kHz filter enabled.
Equipment Setup
Connect equipment as shown in Figure 2-6
Set DMM to ACV, autorange
Test Procedure
1. Reset the AFG:
*RST;*CLS Reset AFG and clear
status registers
2.
Set up the AFG to output a 24 dBm sinewave with the 250 kHz filter
enabled:
VOLT 24DBM; Set amplitude
:OUTP:LOAD 50 50 ohm load
OUTP:FILT:FREQ 250KHZ 250 kHz filter
OUTP:FILT ON Enable filter
INIT:IMM Initiate waveform
Figure 2-6. Equipment Setup for Test 2-5 and Test 2-6
Agilent E1445A Service Manual Verification Tests 49