Agilent Technologies 75000 Series C Portable Generator User Manual


 
Functional Verification: Self-Test
Description
The AFG self-test performs the following internal checks:
internal interrupt lines
waveform select RAM
segment sequence RAM
waveform segment RAM
DDS/NCO operation
sine wave generation
arbitrary waveform generation
marker generation
waveform cycle and arm counters
sweep timer
frequency-shift keying
stop trigger
DC analog parameters (amplitude, offset, attenuators, filters,
calibration DACs)
Test Procedure
1. Remove any connections to the AFG front panel.
2. Reset the AFG:
*RST;*CLS Reset AFG and clear
status registers
3.
Execute the AFG self-test:
*TST? Self-test command
4.
Read the result. A "0" indicates that the test passed. A "1" indicates a
failure. Read the error queue using the SYST:ERR? command until
the error message is "No error".
Agilent E1445A Service Manual Verification Tests 19