Agilent Technologies 75000 Series C Portable Generator User Manual


 
Functional Verification: Start Arm In Test (cont’d)
Test Procedure
(cont’d)
4. Set up the AFG to output a 1 MHz sinewave, with an external Start
Arm source:
FREQ 1E6; Set freq to 1 MHz
:VOLT 4VPP Set AFG amplitude
ARM:LAY2:SOUR EXT External Start Arm
source
INIT:IMM Initiate
5.
Verify that no signal appears on the scope. Send the following
command to the Command Module to provide a Start Arm signal to
the AFG:
OUTP:EXT:LEV 0
6.
Verify that a 1 MHz sinewave appears on the scope.
22 Verification Tests Agilent E1445A Service Manual