Functional Verification: Output Relay Test
Description
The purpose of this test is to check the output relay.
Test Procedure
1. Reset the AFG:
*RST;*CLS Reset AFG and clear
status registers
2.
Set up equipment as shown in Figure 2-4:
3. Set up the AFG to output a 1 MHz sinewave:
FREQ 1E6; Set freq to 1 MHz
:VOLT 4VPP Set AFG amplitude
INIT:IMM Initiate
4.
Verify that a 1 MHz sinewave appears on the scope.
5. Disable the Output relay:
OUTP OFF
6.
Verify that no signal appears on the scope.
Figure 2-4. Output Relay Test Setup