Agilent Technologies 75000 Series C Portable Generator User Manual


 
Test 2-6: AC Flatness - 10 MHz Filter
Description
The purpose of this test is to verify that the AFG meets its specifications for
AC flatness with the 10 MHz filter enabled.
Equipment Setup
Connect equipment as shown in Figure 2-6
Set DMM to ACV, autorange
Test Procedure
1. Reset the AFG:
*RST;*CLS Reset AFG and clear
status registers
2.
Set up the AFG to output a 24 dBm sinewave with the 10 MHz filter
enabled:
VOLT 24DBM; Set amplitude
:OUTP:LOAD 50 50 ohm load
OUTP:FILT:FREQ 10MHZ 10 MHz filter
OUTP:FILT ON Enable filter
INIT:IMM Initiate waveform
3.
Set AFG output to the reference frequency (1 kHz):
FREQ 1000 Set frequency
4.
Measure the amplitude with the DMM, convert the reading to dBm,
and note the reading for future reference:
Reference Level (dBm) = 20 × log ïReading (volts)ï+13.0103
5.
Set the AFG to the crossover frequency (lowest frequency that the
Power Meter can measure):
FREQ 1E5 Set frequency
6.
Measure the amplitude with the DMM and note the reading for future
reference.
52 Verification Tests Agilent E1445A Service Manual