Agilent Technologies 75000 Series C Portable Generator User Manual


 
Functional Verification: Start Arm In Test
Description
The purpose of this test is to check the Start Arm In port. The "TRIG OUT"
port of the Command Module is used to send a Start Arm signal to the AFG.
Test Procedure
1. Reset the AFG:
*RST;*CLS Reset AFG and clear
status registers
2.
Set up equipment as shown in Figure 2-2:
3. Send the following commands to the Command Module to output 0 V
to the "Trig Out" port:
*RST
OUTP:EXT:STAT ON
OUTP:EXT:SOUR INT
OUTP:EXT:LEV 1
Figure 2-2. Start Arm In Test Setup
Agilent E1445A Service Manual Verification Tests 21