Test 2-1: DC Zeros
Description
The purpose of this test is to verify that the AFG meets its specifications for
DCV accuracy for an output of zero volts. An arbitrary waveform
consisting of zeros is used. The amplitude is varied in order to test each
attenuator.
Equipment Setup
Connect equipment as shown in Figure 2-5
Set DMM to: DCV, 100 mV range
Test Procedure
1. Reset the AFG:
*RST;*CLS Reset AFG and clear
status registers
2.
Delete all sequences and segments from memory:
LIST:SSEQ:DEL:ALL Delete all sequences
LIST:SEGM:DEL:ALL Delete all segments
Figure 2-5. Equipment Setup for Test 2-1 thru Test 2-4