Agilent Technologies 75000 Series C Portable Generator User Manual


 
Test 2-9: Total Harmonic Distortion
Description
The purpose of this test is to verify that the AFG meets its specifications for
sine wave total harmonic distortion (THD).
Equipment Setup
Connect equipment as shown in Figure 2-10
Set Spectrum Analyzer to:
Ref Level = 25 dBm
Freq Span = 1 kHz
Resolution BW = 30 Hz
Video BW = 30 Hz
NOTE
These are recommended settings only. Adjust your Spectrum Analyzer as
necessary.
Figure 2-10. Equipment Setup for Test 2-9 and Test 2-10
64 Verification Tests Agilent E1445A Service Manual