Command Descriptions
2–76
TDS Family Oscilloscope Programmer Manual
?
PROBECal
CH
:<x>
FAIl signifies that probe cal has failed for the selected channel.
INIT signifies the probe cal has not yet been run for the selected channel
PASS signifies that probe cal has passed for the selected channel.
CH2:PROBECAL?
might return PASS indicating that probe cal has passed for channel 2.
CH<x>:PROBEFunc:EXTAtten:
TDS 500C & 700C Only
Sets and queries the external attenuation for the specified channel to the specified
value. This command deals with the attenuation factor rather than the gain factor.
This is unlike the CH<x>:PROBE? command.
Vertical
CH<x>:PROBEFunc:EXTAtten:<NR3>
PROBEFunc
CH
:<x>
EXTAtten
:
?
<Space>
<NR3>
<NR3> is the attenuation value specified as a multiplier in the range 1.00E–6 to
1.00E+6. The default is 1.00.
CH2:PROBEFUNC:EXTATTEN 15.00E+0
sets the external attenuation to 15X.
CH<x>:PROBEFunc:EXTDBatten:
TDS 500C & 700C Only
Sets and queries the external attenuation for the specified channel to the specified
value in dB. Note that 1X = 0 dB, 10X = 20 dB, 100X = 40 dB, etc.
Vertical
Outputs
Examples
Group
Syntax
Arguments
Examples
Group