Tektronix TDS 420A Sprinkler User Manual


 
Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–157
HIStogram:COUNt
TDS 500C & 700C Only
Zeros the counts for histograms. If histograms are on then the counts start
counting up again.
Histogram
HIStogram:COUNt RESET
:
COUNt
HIStogram
RESET
<Space>
RESET indicates the need to zero the count.
HISTOGRAM:COUNT RESET
would zero the count.
HIStogram:DISplay
TDS 500C & 700C Only
Selects the way the histogram is displayed, if at all.
Histogram
HIStogram:DISplay { LINEAr | LOG | OFF }
HORizontal:DISplay?
:
DISplay
HIStogram
?
<Space>
OFF
LOG
LINEAr
LINEAr to display the count in each bin. This is the default value.
LOG to display the log of the count in each bin. Log scaling provides better visual
detail for bins with low count.
Group
Syntax
Arguments
Examples
Group
Syntax
Arguments