ResMed AutoSet CS 2 Portable Generator User Manual


 
43
TECHNICAL SPECIFICATIONS
ENGLISH
TECHNICAL SPECIFICATIONS
DIMENSIONS (H X W X D)
142 x 248 x 293mm (5.6" x 9.8" x 11.5")
WEIGHT
Flow generator with cover: 3.7kg
Flow generator with H
UMIDAIRE 2I (empty): 4.2kg
ENVIRONMENTAL CONDITIONS
Operating temperature: +5
o
C to +40
o
C
Storage and transport temperature: -20
o
C to +60
o
C
Humidity — operating, storage and transportation: 10%–95% non-condensing
HOUSING CONSTRUCTION
Injection moulded plastic
AIR TUBING
1 x 2m
AIR FILTER
Synthetic fibre
PATIENT CONNECTION PORT
22mm taper, compatible with EN 1281-1:1997 Anaesthetic & Respiratory Equipment
- Conical Connectors
POWER SUPPLY
AC Input range: 110–120 V and 220–240 V; 50–60 Hz; 60 VA
DC: 30V (via ResMed DC/DC converter)
PERFORMANCE
Operating pressure range: 4–20 cmH
2
O
SOUND PRESSURE LEVEL
<30 dB (tested in accordance with the requirements of ISO 17510-1)
BATTERY TYPE
Lithium
ALARM SPECIFICATIONS
All alarms comply with EN 475 for Medium priority alarms: 3 beeps (935 Hz) repeated
at intervals of 25 seconds; yellow LED flashes.
HIGH MAXIMUM MASK PRESSURE ALARM
Stops air delivery at 25 cmH
2
O >700 msec.
HIGH MEAN MASK PRESSURE ALARM
Stops air delivery when mean mask pressure exceeds 15 cmH
2
O (averaged over
approx 1 minute).
LOW MASK PRESSURE ALARM
Mask pressure is less than 3 cmH
2
O for longer than 5 seconds while the patient is
using the A
UTOSET CS2.