SETTINGS MENU
You will only see the options relevant to your therapy. In CPAP (Continuous Positive
Airway Pressure) mode, the clinician sets a pressure that will be constant throughout
treatment. In ASV-CS mode, A
UTOSET CS2 continually measures your breathing and
regulates the air pressure accordingly.
MASK
Select the type of mask that you will be using (you will only need to adjust this if you
have more than one mask). The following table shows the setting that should be
selected for each mask type. If your mask type is not mentioned below, please contact
your clinician. Not all mask types can be used with the A
UTOSET CS2.
Table 2: Selecting mask options in the Settings menu
LEARN CIRCUIT
The Air Circuit Learn function takes approximately 20 seconds to run. It allows the unit
to ’learn’ what pressure it needs to build in to its calculations, based on the
components in the system. Select LEARN CIRCUIT every time you add or remove a
component (eg mask, humidifier) or move to a significantly different altitude.
To run this function:
1. Set up the entire A
UTOSET CS2 system (flow generator, mask, humidifier etc) as
you wish to use it.
2. Make sure that the mask is unobstructed so air can flow from the mask to the flow
generator.
3. Turn on the A
UTOSET CS2 at the power switch.
4. Move through the menus until you reach LEARN CIRCUIT and select ’yes’ by
pressing the Left key.
5. A message will remind you to check that the airflow from the mask is clear. Select
’start’ with the Left key.
6. A progress screen is displayed while the Air Circuit Learn function is running.
Option in Settings menu Use setting with:
Vista VISTA
Ultra U
LTRA MIRAGE
Full Face M
IRAGE™ FULL FACE MASK SERIES II
U
LTRA MIRAGE FULL FACE MASK
Activa ACTIVA
Papillon Papillon
LEARNING CIRCUIT