ResMed AutoSet CS 2 Portable Generator User Manual


 
39
TROUBLESHOOTING
ENGLISH
TROUBLESHOOTING
If there is a problem, try the following suggestions. If the problem cannot be solved,
contact your equipment supplier or ResMed. Do not attempt to open the flow
generator.
Problem Possible cause Solution
Insufficient air delivered from
flow generator
Air filter is dirty. Replace air filter.
Air tubing is kinked or
punctured.
Straighten or replace tubing.
Flow generator operating
erratically, ie, switching on and
off, pressure varying at random
Water may have entered the
air tubing.
Remove the water from the
air tubing.
Pressure rises inappropriately You are talking, coughing,
voluntarily breath-holding (eg
while rolling over in bed), or
intentionally breathing in an
unusual manner.
Avoid talking with the mask
on, and breathe as normally
as possible.
Severe mask leak or mouth
leak (greater than 30 L/min).
Correct your leak condition.
Alarm stays on continuously Blockage in air circuit. Ensure there is no barrier to
the passage of air.
Internal malfunction. Return the device to your
equipment supplier for
service.
No LCD display Power not connected or
switch at back is not on.
Ensure the power cable is
connected, and that the
switch at the back of the unit
is in the “on” position.
FAULT message displayed on
LCD
Internal malfunction. Return the device to your
equipment supplier for
service.
System Error 7 message Device power-on has been
interrupted by breathing into
the mask during warm up.
Turn the flow generator off
and on again. Do not place
mask on face during warm up
(30 seconds). Select LEARN
CIRCUIT (page 26). If the
error is not cleared, contact
your equipment supplier for
service.
Internal malfunction. Return the device to your
equipment supplier for
service.
eng.book Page 39 Friday, November 3, 2006 2:43 PM