Anritsu MG3641A Portable Generator User Manual


 
SECTION 7 PERFORMANCE TEST
7-8
7.3.4 FM deviation and FM distortion
(1) Test specifications
Range 0 to 125 Hz (125 kHz, <250 kHz)
0 to 250 Hz (250 kHz, <500 kHz)
0 to 500 Hz (500 kHz, <1 MHz)
0 to 1 kHz (1 MHz, <2 MHz)
0 to 2 kHz (2 MHz, <4 MHz)
0 to 4 kHz (4 MHz, <8 MHz)
0 to 10 kHz (8 MHz, <16 MHz)
0 to 25.6 kHz (16 MHz, <32 MHz)
0 to 51.2 kHz (32 MHz, <64 MHz)
0 to 102 kHz (64 MHz, <128 MHz)
0 to 256 kHz (128 MHz, <256 MHz)
0 to 512 kHz (256 MHz, <512 MHz)
0 to 1024 kHz (512 MHz, 1040 MHz)
0 to 2048 kHz (>1040 MHz, MG3642A only)
Accuracy with reference to Source=Int1 1 kHz, and in a 300 Hz to 3 kHz demodulation band:
± (5 % of set value +10 Hz) (0.4 MHz, <512 MHz)
± (5 % of set value +20 Hz) (512 MHz, 1040 MHz)
± (5 % of set value +40 Hz) (>1040 MHz, MG3642A)
Distortion with reference to 16 MHz, and Source=Int1 1 kHz
<40 dB (FM=3.5 kHz deviation)
<45 dB (FM=22.5 kHz deviation)
(2) Measuring instrument for test
Modulation analyzer
Distortion meter
(3) Setup
Fig. 7-4. FM Deviation and FM Distortion
RF
IN
AF
OUT
INPUT
Modulation analyzer
Distortion meter
Demodulation
signal
MG3641A/MG3642A