Anritsu MG3641A Portable Generator User Manual


 
SECTION 6 GPIB
6-51
6.8 IEEE488.2 Common Command
*IDN? Identification Query
Function Reads out the manufacturer name, model number etc. of the product.
Response ANRITSU, MG3641A, 0, 1 or ANRITSU, MG3642A, 0, 1
*OPC Operation Complete Command
Function Set the bit-0 digit in the standard event status register when all the pending selected-device
operations have been completed.
Parameter None
*OPC? Operation Complete Query
Function Sets 1 in the output queue to generate a MAV summary message when all the pending selected-
device operations have been completed.
Response 1
*TST? Self-test Query
Function Executes a self-test and returns the results of any errors.
Response Result = 0 : Indicates that the self-test completed without errors.
1 : Indicates that the self-test is not completed, or is completed with errors.
*WAI Wait-to-continue Command
Function Keeps the next command on stand-by state while the device is executing a command.
Parameter None
*CLS Clear Status Command
Function Clear the status byte register.
Parameter None
*ESE Standard Event Status Enable Command
Function Sets or clear the standard event status enable register.
Parameter Register = 0 to 255
*ESE? Standard Event Status Enable Query
Function Reads out the standard event status enable register.
Response Register = 0 to 255
*ESR? Standard Event Status Register Query
Function Reads out the current value in the standard event status register.
Response Register = 0 to 255
*SRE Service Request Enable Command
Function Sets the bits in the service request enable register.
Parameter Register = 0 to 255
*SRE? Service Request Enable Query
Function Reads out the service request enable register.