Agilent Technologies e4419b Yard Vacuum User Manual


 
Specifications
Power Meter Supplemental Characteristics
1-10 Agilent E4418B/E4419B Service Guide
For Agilent E-series power sensors
In FAST mode, within the range -50 dBm to +17 dBm, for a 10 dB
decreasing power step, the settling time is 10 ms
1
for the Agilent E4418B
and 20 ms
1
for the Agilent E4419B
1
When a decreasing power step crosses the power sensor’s auto-range
switch point, add 25 ms. Refer to the relevant power sensor manual for
further information.
For Agilent E-series power sensors in normal and x2 speed modes, manual
filter, 10 dB decreasing power step:
Table 1-5: Settling Time
Number of
Averages
12481632641282565121024
Settling Time (s)
(Normal Mode)
0.07 0.12 0.21 0.4 1 1.8 3.3 6.5 13 27 57
Response Time (s)
(x2 mode)
0.04 0.07 0.12 0.21 0.4 1 1.8 3.4 6.8 14.2 32