Agilent Technologies e4419b Yard Vacuum User Manual


 
Measurement Uncertainty Analysis – Power Reference Level Test
Measurement Introduction
Agilent E4416A/E4417A Service Guide B-9
Table B-1: Measurement Repeatability Results
Measurement ADJUST REF. REF. LEVEL TEST
1 0.999548951 0.999499495
2 0.999541333 0.999554257
3 0.999449078 0.999414037
4 0.999546994 0.999509401
5
1
0.999138526 0.9991726
6 0.999647055 0.999454911
7 0.999539391 0.999515949
8 0.999567191 0.999558795
9 0.999600963 0.999554633
10 0.999405584 0.999377721
11 0.999523392 0.999566084
12 0.999504386 0.999504323
13 0.999443652 0.999463611
14 0.999513893 0.999553753
15 0.999589288 0.999496811
16 0.999485404 0.999529486
17 0.999559381 0.999534985
18 0.999524217 0.99948097
19 0.999523013 0.999518899
20 0.999534276 0.999488886
Max 0.999647 0.999566084