Agilent Technologies e4419b Yard Vacuum User Manual


 
Troubleshooting
Introduction
6-10 Agilent E4418B/E4419B Service Guide
Figure 6-4: Troubleshooting Guide - RS232/422 Interface
Start
Power up
meter
POST
Pass?
Which POST Test
failed?
Select
System/Inputs
Remote Interface
Select Interface
RS422
Connect RS232 Self Test
Connector to RS232/422
Connector on rear panel
Select
System/Inputs
More
Service
Self Test
Individual
More
Serial Interface
RS422 Loop Back
Run Test
Select
Done
System/Inputs
Remote Interface
Select Interface
RS232
Connect RS422 Self Test
Connector to RS232/422
Connector on rear panel
Select
System/Inputs
More
Service
Self Test
Individual
More
Serial Interface
RS232 Loop Back
Run Test
Test Pass?
Connect power meter
to a GP-IB controller
and send the
command “*RST”
Did power
meter reset?
Select
System/Inputs
Remote Interface
Select Interface
GPIB (IEEE 488)
Prev
Command Set
SCPI
No fault found with
RS232/422 serial
interface check user
application/
implementation
Test Pa ss?
Investigate power
meter failure
A2 Processor assy
defective
A4 Comms assy
defective
End
Yes
Yes
Yes
NoYes
No
Serial Interface
All/Other
No
No