Remote Operation 73
Table 5-1. Remote Programming Ranges (continued)
Function
HPSL Command
(Short Form) Range of Values
Constant Resistance (CR)
Set Range "RES:RANG value"
6060B 6063B
Low Range
≥
0 or
≤
1
Ω≥
0 and
≤
24
Ω
Middle Range
>1
Ω
and
≤
k
Ω≤
24
Ω
and
≤
24 k
Ω
High Range
>1 k
Ω
and
≤
k
Ω
>24 k
Ω
and
≤
240 k
Ω
Set Main Level ’RES value"
Low Range
0 to 1
Ω
0 to 24
Ω
Middle Range
1
Ω
to 1 k
Ω
24
Ω
to 24 k
Ω
High Range
10
Ω
to 10 k
Ω
240
Ω
to 240 k
Ω
Set Slew Rate
Low Range "VOLT: SLEW value" Same as CV slew rate
Middle/High Range "CURR:SLEW value" Same as CC slew rate
Set Transient Level "RES:TLEV value" Same as main CR level
*Set Triggered Level "RES:TRIG value" Same as main CR level
Constant Voltage (CV) 6060B 6063B
Set Main Level "VOLT value" 0 to 60 V 0 to 240 V
Set Slew Rate "VOLT:SLEW value" 1000 to 5,000,000 V/s 4000 to 2,000,000 V/s
Set Transient Level "VOLT:TLEV value" Same as main CV level
*Set Triggered Level "VOLT:TRIG value" Same as main CV level
Transient Operation
Set Frequency "TRAN:FREQ value" 0.25 to 10000 Hz
Set Duty Cycle "TRAN:DCYC value" 3 to 97% (0.25 Hz to 1 kHz
6-94% (1 kHz-10 kHz)
*Set Pulse Width "TRAN:TWID value" 0.00005 to 4 seconds
Current Protection 6060B 6063B
*Set Current Level "CURR:PROT value" 0 to 61.2 A 0 to 10.2 A
*Set Delay Time "CURR:PROT:DEL value" 0 to 60 seconds
*Can only be programmed remotely via the GPIB.