Local Operation 59
4. Set Transient Level-The transient resistance level "R:TLV" is meaningful only if transient operation (described
later) is turned on.
a. Set the transient level to 0.8 ohm by pressing
. Remember that in the low range the transient level must be set higher than
the main level.
b. Press
again and note that the display indicates "R:TLV 0.8000". Note that you can use the Input
ENTRY keys to increment and decrement the transient resistance level. Operation is similar to that described
for the main resistance level.
Setting CV Values
The CV values for the are programmed by pressing the applicable FUNCTION keys and setting the desired values using the
ENTRY keys. The display identifies the selected function; for example "V:TLV" identifies the transient voltage level.
Range
The voltage values can only be programmed in one range. The valid CV values are listed in Table 4-4 along with the
applicable front panel key and display identifier. All voltage levels are programmed in volts and the voltage slew rate is
programmed in volts/microsecond.
Table 4-4. CV Programming Ranges
Function Key Display Range of Values
6060B 6063B
Set Main Level "VOLT value" 0.000 to 60.000 0.000 to 240.000
Set Slew Rate
(shifted)
"V:SLW value"
0.0010 to 0.5000 (V/
µ
s)
(Note 1)
0.0040 to 2.000 (V/
µ
s)
Set Transient Level "V:TLV value" 0.000 to 60.000 (Note 2) 0.000 to 240.00
Notes:
1. There are 12-discrete steps within the voltage slew range. Because of bandwidth limitations, only 9 slew rate steps can be
achieved (see Table 1-1). Any slew rate can be programmed. (There are no upper and lower limits that would cause an error.)
The Electronic Load automatically selects one of the 12 slew rates that is closest to the programmed value. See Chapter 2
Operation Overview in this manual.
2. The transient voltage level is meaningful only if transient operation is turned on. The transient voltage level must be set
to a higher value than the main voltage level. See Transient Operation.
Examples
The following examples illustrate how to program CV values. Before you do these examples, press
to set the CV values to their factory default values.
1. Set Main level
a. Set the main voltage level to 20 volts by pressing
b. Press
again
and check that the display indicates "VOLT 20.000".