Agilent Technologies 5951-2826 Tiller User Manual


 
72 Remote Operation
80 OUTPUT 705;"TRAN ON
90 OUTPUT 705;"INPUT ON"
.
.
.
.
200 OUTPUT 705;"*TRG"
210 END
Line 10: Turns off Electronic Load input.
Line 20: Selects the CR mode.
Line 30: Selects the main resistance level to 10 ohms.
Line 40: Sets the transient resistance level to 5 ohms. Remember in the 1 to 1k range, the transient resistance level
must be set to a lower level than the main resistance level.
Line 50: Selects the GPIB as the trigger source.
Line 60: Sets the current slew rate to the fastest rate. Remember that in the middle range, the resistance slew rate
is programmed in amps/second.
Line 70: Selects pulsed transient operation and sets the pulse width to 1 millisecond.
Line 80: Turns on the transient generator.
Line 90 Turns on Electronic Load input.
Line 100
to Other commands are executed.
Line 190
Line 200: The *TRG command generates a 1 millisecond pulse at the Electronic load input.
Table 5-1. Remote Programming Ranges
Function
HPSL Command
(Short Form) Range of Values
Constant Current (CC)
Set Range "CURR:RANG value"
6060B 6063B
Low Range
0 and
6 A
0 and
1 A
High Range
> 6 and
60 A > 1 and
10 A
Set Main Level ’CURR value"
Low Range 0 to 6 A 0 to l A
High Range 0 to 60 A 0 to 10 A
Set Slew Rate "CURR:SLEW value"
Low Range . 100 to 500,000 A/s 1.7 to 83,000 A/s
High Range 1000 to 5,000,000 A/s 17 to 830,000 A/s
Set Transient Level "CURR:TLEV value" Same as CC main level
*Set Triggered Level "CURR:TRIG value" Same as CC main level