Agilent Technologies 5951-2826 Tiller User Manual


 
General Information 15
Table 1-2. Supplemental Characteristics (continued)
Fuse: The ac input is protected by a fuse located in a module on the rear panel; 0.5AM for l00/120 Vac
input; 0.25AM for 220/240 Vac input.
Maximum VA: 60
Peak Inrush Current: 2.5 A (typical)
PROGRAMMABLE SLEW RATE: (For any given input transition, the time required will be either the total slew time
or a minimum transition time, whichever is larger. The minimum transition time increases when operating with input
currents under 1 AM (6060B) or 0.2 AM (6063B) and decreases with input currents over 20 A (6060B) or 2 A (6063B).
The following are typical values;
±
25% tolerance.)
Current Slew Rate:
Model 6060B (Ac performance specified from 3 to 60 V)
Rate # High Range Step Low Range Step Transition Time
1 1 A/ms 0.1 A/ms 8.0 ms
2 2.5 A/ms 0.25 A/ms 3.2 ms
3 5 A/ms 0.5 A/ms 1.6 ms
4 10 A/ms 1 A/ms 800
µ
s
5 25 A/ms 2.5 A/ms 320
µ
s
6 50 A/ms 5 A/ms 160
µ
s
7 0.1 A/
µ
s 10 A/ms 80
µ
s
8 0.25 A/
µ
s 25 A/ms 32
µ
s
9 0.5 A/
µ
s 50 A/ms 16
µ
s
10 1 A/
µ
s 0.1 A/
µ
s 12
µ
s
11 2.5 A/
µ
s 0.25 A/
µ
s 12
µ
s
12 5 A/
µ
s 0.5 A/
µ
s 12
µ
s
Model 6063B (Ac performance specified from 3 to 240 V)
Rate # High Range Step Low Range Step Transition Time
1 0.17 A/ms 17 A/s 8.0 ms
2 0.42 A/ms 42 A/s 3.2 ms
3 0.83 A/ms 83 A/s 1.6 ms
4 1.7 A/ms 0.17 A/ms 800
µ
s
5 4.2 A/ms 0.42 A/ms 320
µ
s
6 8.3 A/ms 0.83 A/ms 160
µ
s
7 17 A/ms 1.7 A/ms 80
µ
s
8 42 A/ms 4.2 A/ms 32 µs
9 83 A/ms 8.3 A/ms 20 µs
10 0.17 A/µs 17 A/ms 20 µs
11 0.42 A/µs 42 A/ms 16 µs
12 0.83 A/µs 83 A/ms 16 µs