Making Measurements
Making overshoot and preshoot measurements
Overshoot
Overshoot is distortion that follows a major edge transition expressed as a
percentage of Amplitude. The X cursors show which edge is being measured
(edge closest to the trigger reference point).
Automatic Overshoot measurement
Rising edge overshoot =
local Maximum Top
Amplitude
-----------------------------------------------------
X 100
Falling edge overshoot =
Base local Minimum
Amplitude
------------------------------------------------------
X 100
local Maximum
Overshoot
Overshoot
local Minimum
Top
Base