Specifications
86 AFG3000 Series Quick Start User Manual
Output Characteristics
Sine Wave
✔Flatness
(at 1.0 V
p-p
amplitude (+4 dBm),
relative to 100 kHz)
<5 MHz: ±0.15 dB
≥5 MHz to 10 MHz: ±0.3 dB
✔Harmonic Distortion
(at 1.0 V
p-p
amplitude)
10 Hz to 20 kHz: <–60 dBc
≥20 kHz to 1 MHz: <–55 dBc
≥1 MHz to 10 MHz: <–45 dBc
✔Total Harmonic Distortion
(at 1 V
p-p
amplitude)
10 Hz to 20 kHz: <0.2%
✔Spurious
1
(nonharmonic)
(at 1 V
p-p
amplitude)
10 Hz to 1 MHz: <–60 dBc
≥1 MHz to 10 MHz: <–50 dBc
Phase Noise, typical
(at 1 V
p-p
amplitude)
10 MHz: <–110 dBc/Hz at 10 kHz offset
Residual Clock Noise, typical –63 dBm
Square Wave
✔Rise Time/Fall Time
2
(Amplitude: ≤ 10 V
p-p
)
≤ 50 ns
Jitter (rms), typical 500 ps
Pulse
Pulse Width 80 ns to 999.99 s
Resolution 10 ps or 5 digits
Pulse Duty 0.001% to 99.999%
Leading Edge/Trailing Edge 50 ns to 0.625 * Pulse Period
Resolution 10 ps or 4 digits
Overshoot, typical <5%
Jitter (rms), typical 500 ps
Ramp
Linearity
3
, typical ≤ 0.2% of peak output
Symmetry 0% to 100.0%
Noise
Bandwidth, typical 10 MHz
Arbitrary
Rise Time/Fall Time, typical ≤ 80 ns
Jitter (rms), typical 4 ns
1. Exclude harmonics and channel crosstalk.
2. ≤ 50 ns (Amplitude: >10 V
p-p
) (Typical)
3. Frequency: 1 kHz, Amplitude: 1 V
p-p
, Symmetry: 100%
10% to 90% of amplitude range