Section 2.0 – Installation
2.1 Unpacking and Inspection
The Clarity II Turbidimeter is a complete system for the determination of turbidity in drinking water.
The system consists of the analyzer, sensor(s), cable(s), and flow chamber/debubbler(s). Consult the
table to verify that you have received the parts for the option you ordered.
2.2 Installation
2.2.1 General Information
1. Although the analyzer is suitable for outdoor use, do not install it in direct sunlight or in
areas of extreme temperatures.
2. Install the analyzer in an area where vibration and electromagnetic and radio frequency
interference are minimized or absent.
3. Keep the analyzer and sensor wiring at least one foot from high voltage conductors. Be sure
there is easy access to the analyzer.
4. The analyzer is suitable for panel, pipe, or surface mounting. Refer to the table below.
Installation 5
Item Model/part number
Single Input Turbidity Analyzer with HART 56-03-27-38-HT
Dual Input Turbidity Analyzer with HART 56-03-27-37-HT
Sensor-EPA standard 8-0108-0002-EPA
Sensor-ISO standard 8-0108-0003-ISO
Cable-3 ft (0.9 m) 24138-00
Cable-20 ft (6.1 m) 24097-00
Cable-50 ft (15.2 m) 24098-00
Calibration cup 24101-00
Molded chamber/debubbler 24170-00
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with the National Electrical Code
(ANSI/NFPA-70) and/or any other applicable national or local codes.
Type of Mounting Figure
Panel 2-1
Wall and Pipe 2-2
WARNING