Agilent Technologies B Weather Radio User Manual


 
179
Chapter 3: Reference
Error Messages
No more Pattern resources available for this
pod pair
This error occurs when you have used up all the pattern resources
available. Each pod pair has about 28 pattern resources. Some pattern
events use more than 1 resource.
Possible Solutions
Keep labels within a pod pair
If a label (bus) spans pod pairs (for example, pods 2 and 3) then when you
use the label in a trigger sequence it will use up at least one pattern
resource on both pod pairs. If you hook up your probes in such a way that
the signals are on a single pod pair, you can free up a pattern resource on
the other pod pair.
Move some labels to another pod pair
Each pod pair has its own set of pattern resources. Putting your two most-
used labels on different pod pairs can improve your resource usage.
You may find more information in the Branch expression is too
complex (see page 171) error message help for the 800 and 1250 Mb/s
modes.
No Trigger action found in the trace
specification
This warning occurs when the trigger sequence you specified does not
have at least one trigger and fill memory or trigger and goto action.
The analyzer will still acquire data, but you will need to manually stop
it. For example, when you use the Run until user stop function, to
avoid popping up the status window, set the popup level to errors only.