Agilent Technologies 86100b Sprinkler User Manual


 
1-27
Setting Up the Infiniium DCA
Step 12. Perform a Module Calibration
To perform a module calibration
1 On the Calibrate menu, choose All Calibrations.
The All Calibrations dialog box opens. This dialog box allows you to view
the calibration status of modules and initiate the calibration.
2 Touch the Modules tab.
The Modules tab opens and allows you to perform a module calibration on
either the left or right module.
3 Touch the Calibrate Module button (located at the bottom of
the tab) for the module you need to calibrate. This starts the
Module calibration.
4 Remove all external devices and signals from the module.
The module calibration uses known signal levels in the instrument. Introduc-
ing an outside signal source will interfere with the calibration factors and
decrease the accuracy of the calibration.
5 Touch Continue. A progress indicator appears.
6 After the calibration completes, choose All Calibrations on the Calibration
menu.
7 The Module tab shows the status of the calibration. The tab will look similar to
the following figure. Notice that the tab indicates that the Current user