Agilent Technologies 86100b Sprinkler User Manual


 
1-26
Setting Up the Infiniium DCA
Step 12. Perform a Module Calibration
Step 12. Perform a Module Calibration
A module calibration establishes calibration factors for all installed modules.
The calibration factors compensate for imperfections in the measurement sys-
tem (for example, variations due to ambient temperature), resulting in the
best instrument precision. It is recommended that you perform this calibration
routinely for best measurement accuracy.
The module calibration is recommended whenever the following occurs:
Instrument power has been cycled
A module has been removed and then reinserted
A change in the temperature of the mainframe exceeds 5°C compared to the
temperature of the last vertical (amplitude) calibration (T > 5°C)
The time since the last calibration has exceeded 10 hours
Refer to the Help for more information about instrument calibrations.
CAUTION Electrical channel input circuits and the trigger input circuit can be damaged
by electrostatic discharge (ESD). Therefore, avoid applying static discharges
to the front-panel input connectors. Prior to connecting any coaxial cable to
the connectors, momentarily short the center and outer conductors of the
cable together. Avoid touching the front-panel input connectors without first
touching the frame of the instrument. Be sure that the instrument is properly
earth-grounded to prevent buildup of static charge. Wear a wrist-strap or heel-
strap.
Warm-up time
Allow the instrument, with modules installed, to warm up for a period of at least one
hour before performing a vertical calibration. This ensures that the module reaches its
equilibrium temperature.