Agilent Technologies 86100b Sprinkler User Manual


 
1-13
Setting Up the Infiniium DCA
Step 5. Connect the Line Cord
only a standby switch and is not a LINE switch. Alternatively, an externally
installed switch or circuit breaker (which is readily identifiable and is easily
reached by the operator) may be used as a disconnecting device.
CAUTION Always use the three-prong AC power cord supplied with this instrument.
Failure to ensure adequate earth grounding by not using this cord may cause
instrument damage.
CAUTION Do not connect ac power until you have verified the line voltage is correct as
described in Table 1-7. Damage to the equipment could result.
CAUTION This instrument has autoranging line voltage input. Be sure the supply voltage
is within the specified range.
Table 1-7. Line Power Requirements
Power 115 VAC: 110 VA MAX. / 60 WATTS MAX. / 1.1 A MAX.
230 VAC: 150 VA MAX. / 70 WATTS MAX. / 0.6 A MAX.
Voltage nominal: 115 VAC / 230 VAC CAT II
range 115 VAC: 90132 V
range 230 VAC: 198254 V
Frequency nominals: 50 Hz / 60 Hz
range: 4763 Hz
Table 1-8. Available Line Cords
Plug Type Cable Part No. Plug Description
Length
(in/cm)
Color Country
250V 8120-1351
8120-1703
Straight *BS1363A
90°
90/228
90/228
Gray
Mint Gray
United Kingdom, Cyprus,
Nigeria, Zimbabwe,
Singapore
250V 8120-1369
8120-0696
Straight *NZSS198/ASC
90°
79/200
87/221
Gray
Mint Gray
Australia, New Zealand