Calibration 85
Explanation
LINE 10-20 Specify select code, address, and channel (default 705, 1)
LINE 30 Turn calibration mode on
LINE 40-90 Assign variables for subprograms (see module calibration tables)
LINE 100 Store new constants in EEROM when calibration complete
LINE 110 Turn calibration mode off
LINE 140 Current calibration subroutine
LINE 200-220 Select channel, current mode, and range
LINE 230 Set high calibration point
LINE 240 If high current range, wait for internal current shunt to stabilize
LINE 260 Send measurement in amperes for high main calibration point
LINE 270 If high current range, send measurement in amperes for high readback cal point
LINE 280 Set low calibration point
LINE 300 Send measurement in amperes for low main calibration point
LINE 310 If high current range, send measurement in amperes for low readback cal point
LINE 350 Set low calibration point
LINE 360-370 Select transient toggle mode and GPIB trigger source
LINE 380-390 Turn transient mode on and set transient calibration point
LINE 400 Trigger transient level
LINE 410 If high current range, wait for internal current shunt to stabilize
LINE 430 Send measurement in amperes for high transient calibration point
LINE 440 Turn transient mode off
LINE 500 Voltage calibration subroutine
LINE 550-560 Select channel and voltage mode
LINE 570 Set high calibration point
LINE 590 Send measurement in volts for high main calibration point
LINE 600 Send measurement in volts for high readback calibration point
LINE 610 Set low calibration point
LINE 630 Send measurement in volts for low main calibration point
LINE 640 Send measurement in volts for low readback calibration point
LINE 680 Set low calibration point
LINE 690-700 Select transient toggle mode and GPIB trigger source
LINE 710-720 Turn transient mode on and set transient calibration point
LINE 730 Trigger transient level
LINE 750 Send measurement in volts for transient calibration point
LINE 760 Turn transient mode off
LINE 820 Resistance calibration subroutine
LINE 870-890 Select channel, resistance mode, and range
LINE 900 Set high calibration point
LINE 930-940 Calculate and send measurement in ohms for high main calibration point
LINE 950 Set low calibration point
LINE 980-990 Calculate and send measurement in ohms for low main calibration point
LINE 1030-1070 If middle and high range, set high calibration point; otherwise set low point
LINE 1080-1090 Select transient toggle mode and GPIB trigger source
LINE 1100 Turn transient mode on
LINE 1110-1150 If middle and high range, set lower transient point; otherwise set higher point
LINE 1160 Trigger transient level
LINE 1190-1200 Calculate and send measurement in ohms for transient calibration point
LINE 1210 Turn transient mode off