Agilent Technologies 5951-2827 Tiller User Manual


 
12 General Information
Table 1-1. Specifications (continued)
TRANSIENT CURRENT LEVEL
Ranges
Low Range: 0 to 6 A 0 to 1 A
High Range: 0 to 60 A 0 to 10 A
Accuracy
Low Range: ± 0.1% ± 80 mA ± 0.18% ± 13 mA
High Range: ± 0.1% ± 350 mA ± 0.18% ± 50 mA
6060B 6063B
TRANSIENT RESISTANCE LEVEL
Ranges
Low Range: 0.033 to 1
0.20 to 24
Middle Range: 1 to 1000
24 to 10,000
High Range: 10 to 10,000
240 to 50,000
Accuracy
Low Range:
±
0.8% ± 10 m
± 0.8% ± 200 m
with
6 A at input with
1 A at input
Middle Range: ± 0.3% ± 10 mS ± 0.3% ± 0.5 mS
with
6 V at input with
24 V at input
High Range: ± 0.3% ± 10 mS ± 0.3% ± 0.4 mS
with
6 V at input with
24 V at input
TRANSIENT VOLTAGE LEVEL
Range: 0 to 60 V 0 to 240 V
Accuracy: 0.1% ± 300 mV ± 0.15% ± 1.1 V
CURRENT READBACK
*Accuracy: 0.05% ± 65 mA ± 0.12% ± l0mA
*after 30 second wait.
VOLTAGE READBACK
Accuracy: 0.05% ± 45 mV ± 0.1%150 mV
POWER READBACK
Accuracy ±0.2% ± 4 W ±0.2%
±
3 W
PARD (20 Hz to 10 MHz noise)
Current: 4 mA rms/40 mA p-p 1 mA rms/10 mA p-p
Voltage: 6 mV rms 6 mV rms