Agilent Technologies 5951-2827 Tiller User Manual


 
Local Operation 63
Table 4-6. Factory Default Settings
Function Setting
6060B 6063B
Input on/off on on
Short on/off off off
CURR level 0 A 0 A
CURR transient level 0 A 0 A
CURR slew rate
1 A/
µ
s 0.83 A/
µ
s
CURR range 60 A 10 A
*CURR protection level 61.2 A 10.2 A
*CURR protection delay 15 s 15 s
*CURR protection on/off off off
VOLT level 60 V 240 V
VOLT transient level 60 V 240 V
VOLT slew rate
5 V/
µ
s 2 V/
µ
s
RES level
1000
2400
RES transient level
1000
24000
RES range
1000
24000
Mode CC CC
Transient Operation off off
Frequency 1 kHz 1 kHz
Duty Cycle 50% 50%
**Transient mode continuous continuous
*Pulse width 0.5 ms 0.5 ms
*Port0 output signal TTL logic 0 TTL logic 0
*Calibration mode off off
*Trigger source hold hold
*Can only be programmed remotely via the GPIB.
**Continuous transient operation is the only mode of transient operation available
at the front panel. Pulsed, toggled, and continuous transient operating modes may
be programmed remotely via the GPIB.
Changing "Wake-up" Settings
The "wake-up" settings are stored in register 0. At power-on, the Electronic Load will "wake-up" with these values set.
When the Electronic Load is shipped from the factory, its "wake-up" values are the same as its factory default values (see
Table 4-6).
You can change the "wake-up" values to whatever values you wish. You do this by setting them into the Electronic Load
and then saving them in register 0 by pressing
(blue shift key) (shifted Recall key)
.
When power is turned off and on, the Electronic Load will be set to the values you saved in register 0.
The Save 0 operation takes a few seconds to complete. Do not turn power off until the "SAVE 0 "
message goes away indicating that the operation is complete. If you turn off power before completion,
the Electronic Load’s non- volatile memory will be corrupted and the Electronic Load
will need to be recalibrated.