Tektronix TDS 460A Sprinkler User Manual


 
Command Descriptions
2–318
TDS Family Oscilloscope Programmer Manual
<Space>
EITher
?
STAYSHigh
STAYSLow
TRIGger :
MAIn
POLarity
:
PULse
:
:
TIMEOut
STAYSLow (or NEGAtive) indicates that a pulse edge must stay low the required
time period for timeout triggering to occur.
STAYSHigh (or POSITIVe) indicates that a pulse edge must stay high the required
time period for timeout triggering to occur.
EITher indicates either STAYSHigh or STAYSLow polarity.
TRIGGER:MAIN:PULSE:TIMEOUT:POLARITY EITHER
specifies that the polarity of the timeout trigger can be either positive or
negative.
TRIGger:MAIn:PULse:TIMEOut:TIMe
TDS 500C, 600B, & 700C Only
Sets or queries the pulse timeout trigger time, in seconds. This command is
equivalent to selecting Time in the Trigger menu and entering a value with the
keypad or general purpose knob.
Trigger
TRIGger:MAIn:PULse:TIMEOut:TIMe <NR3>
TRIGger:MAIn:PULse:TIMEOut:TIMe?
?
<Space> <NR3>
TRIGger
:
:
PULse
MAIn
TIMEOut
:
:
TIMe
<NR3> is the timeout time period, in seconds.
Arguments
Examples
Group
Syntax
Arguments