Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–189
LOCk { ALL | NONe }
LOCk?
<Space>
?
ALL
NONe
LOCk
ALL disables all front panel controls.
NONe enables all front panel controls. This is equivalent to the UNLock ALL
command.
NOTE. If the digitizing oscilloscope is in the Remote With Lockout State (RWLS),
the LOCk NONe command has no effect. For more information see the ANSI-
IEEE Std. 488.1-1987 Standard Digital Interface for Programmable Instrumen-
tation, section 2.8.3 on RL State Descriptions.
LOCK ALL
locks the front panel controls.
LOCK?
returns NONe when the front panel controls are enabled by this command.
*LRN? (Query Only)
Returns a string listing the digitizing oscilloscope settings, except for configura-
tion information for the calibration values. You can use this string to return the
digitizing oscilloscope to the state it was in when you made the *LRN? query.
Miscellaneous
HEADer, SET?, VERBose
*LRN?
*LRN ?
Syntax
Arguments
Examples
Group
Related Commands
Syntax