Tektronix TDS 460A Sprinkler User Manual


 
Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–291
NANd
NOR
<Space>
?
OR
AND
:
MAIn
TRIGger
LOGIc :: FUNCtion
AND specifies that the instrument will trigger if all the conditions are true.
NANd specifies that the instrument will trigger if any of the conditions are false.
NOR specifies that the instrument will trigger if all of the conditions are false.
OR specifies that the instrument will trigger if any of the conditions are true.
TRIGGER:MAIN:LOGIC:FUNCTION NOR
sets the logical combination of channels to be true when none of the conditions
are true.
TRIGGER:MAIN:LOGIC:FUNCTION?
might return NAND.
TRIGger:MAIn:LOGIc:INPut? (Query Only)
Not on TDS 400A
Returns the main logic trigger input for all channels.
Trigger
TRIGger:MAIn:LOGIc:INPut?
TRIGger :
:
LOGIc
MAIn :
?
INPut
TRIGGER:MAIN:LOGIC:INPUT?
might return :TRIGGER:MAIN:LOGIC:INPUT:CH1 HIGH;CH2 X;CH3 X
Arguments
Examples
Group
Syntax
Examples